Home / Brands / Technomed Europe
Technomed Europe Recalls
3 recalls on record · January 19, 2022 to September 18, 2024 · Medical Devices
Recalls by year
22
24
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
September 18, 2024Medical DevicesFDA
Technomed Europe recalls NEUROSIGN Disposable Stimulating Dissector, REF 5888-00
Hazard: Product has have presence of brown discoloration or substance on the hinge area of the device
September 21, 2022Medical DevicesFDA
Technomed Europe recalls Disposable Subdermal Needle Electrode, Pt/lr, 12 x 0.40mm, product number TE/S46/638, 25 needles per package, sterile.
Hazard: Burns of varying degrees can occur to patients with electrodes left in while in the MRI scanner as a result of inadequate safety information in the IFU.
January 19, 2022Medical DevicesFDA
Technomed Europe recalls Disposable Subdermal Needle Electrode, Corkscrew, Single Use, Sterile; Model Numbers TE/S50715-001-C, TE/S50725-001C, an
Hazard: There is a high risk of the blue hub detaching from the corkscrew needle, resulting the the needle remaining in the patient's scalp.