TELEFLEX MEDICAL Recalls

236 recalls on record · July 25, 2012 to June 1, 2022 · Medical Devices

Recalls by year

12
13
14
15
16
17
18
19
20
21
22

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

February 19, 2020Medical DevicesFDA
TELEFLEX MEDICAL INC recalls HUDSON RCI COMFORT FLO Corrugated Humidification System with Remote Temperature Port, REF 2416, humidifier nebulizer kit
Hazard: There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.
February 19, 2020Medical DevicesFDA
TELEFLEX MEDICAL INC recalls HUDSON RCI COMFORT FLO Corrugated Humidification System, REF 2415, humidifier nebulizer kit
Hazard: There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.
January 22, 2020Medical DevicesFDA
Teleflex Medical recalls Teleflex MEDICAL RuSCH One piece Male External Medium Catheter W/O Tape, REF A1500M
Hazard: The device label incorrectly states that the device is sterile. This product is an externally applied male condom-type catheter designed for the management of male incontinence as an alternative to incontinence pads. In normal use, the device is neither expected nor required to be sterile.
January 22, 2020Medical DevicesFDA
Teleflex Medical recalls Teleflex MEDICAL RuSCH One piece Male External Regular Catheter W/O Tape, REF A1500
Hazard: The device label incorrectly states that the device is sterile. This product is an externally applied male condom-type catheter designed for the management of male incontinence as an alternative to incontinence pads. In normal use, the device is neither expected nor required to be sterile.
December 25, 2019Medical DevicesFDA
Teleflex Medical recalls HUDSON RCI SHERIDAN SHERI-BRONCH Endobronchial Tube REF 5-16135 - Product Usage: The Sheridan Endobronchial Tubes are in
Hazard: The tracheal and brochial swivel connectors are incorrectly labeled.
December 25, 2019Medical DevicesFDA
Teleflex Medical recalls Arrow EZ-IO Intraosseous Vascular Access System 45mm Needle+Stabilizer Kit, Product Code 9079P-VC-005 - Product Usage: T
Hazard: The safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.
December 25, 2019Medical DevicesFDA
Teleflex Medical recalls Arrow EZ-IO Intraosseous Vascular Access System 25mm Needle+Stabilizer Kit, Product Code 9001P-VC-005 - Product Usage: T
Hazard: The safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.
December 25, 2019Medical DevicesFDA
Teleflex Medical recalls Arrow EZ-IO Intraosseous Vascular Access System 15mm Needle+Stabilizer Kit, Product Code 9018P-VC-005 - Product Usage: T
Hazard: The safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.
November 27, 2019Medical DevicesFDA
Teleflex Medical recalls WECK Hem-O-Lok AutoEndo5, 5mm Automatic Endoscopie Hem-o-lok Applier, REF 543965 Automatic Endoscopic Surgical Lig
Hazard: Certain clip appliers have an increase in misloading and/or jamming related complaints that may be due to a discrepancy in the specification a component. Misloading and/or jamming may lead to clips falling out of the applier, clip breakage, or the jaws becoming locked in a partially closed position (jammed).
November 27, 2019Medical DevicesFDA
Teleflex Medical recalls WECK AutoEndo5, Automatic Hem-O-Lok Clip Applier, REF AE05ML Automatic Endoscopic Surgical Ligating Clip Appliers a
Hazard: Certain clip appliers have an increase in misloading and/or jamming related complaints that may be due to a discrepancy in the specification a component. Misloading and/or jamming may lead to clips falling out of the applier, clip breakage, or the jaws becoming locked in a partially closed position (jammed).
September 18, 2019Medical DevicesFDA
Teleflex Medical recalls Sheridan/HVT Novaplus: a) 7.0 mm, REF V5-10314 b) 7.5 mm, REF V5-10315 c) 8.0 mm, REF V5-10316 d) 8.5 mm, REF V5-10
Hazard: Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
September 18, 2019Medical DevicesFDA
Teleflex Medical recalls Sheridan/CF Novaplus 7.0 mm, REF V5-10114 Product Usage: Tracheal tube/airway management
Hazard: Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
September 18, 2019Medical DevicesFDA
Teleflex Medical recalls Hudson RCI Sheridan/HVT: a) 5.0 mm, REF 5-10310 b) 5.5 mm, REF 5-10311 c) 6.0 mm, REF 5-10312 d) 6.5 mm, REF 5-10313
Hazard: Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
September 18, 2019Medical DevicesFDA
Teleflex Medical recalls Hudson RCI Sheridan Uncuffed: a) 2.0 mm, REF 5-10404 b) 2.5 mm, REF 5-10405 c) 3.0 mm, REF 5-10406 d) 3.5 mm, REF
Hazard: Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
September 18, 2019Medical DevicesFDA
Teleflex Medical recalls Hudson RCI Sheridan Preformed: a) 3.0 mm, REF 5-22006, REF 5-22106 b) 3.5 mm, REF 5-22007, REF 5-22107 c) 4.0 mm, REF
Hazard: Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
September 18, 2019Medical DevicesFDA
Teleflex Medical recalls Hudson RCI Sheridan EZ-ENDO: a) 5.0 mm, REF 5-22510 b) 5.5 mm, REF 5-22511 c) 7.0 mm, REF 5-22514 d) 7.5 mm, REF 5-2
Hazard: Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
September 18, 2019Medical DevicesFDA
Teleflex Medical recalls Hudson RCI Sheridan/CF: a) 3.0 mm, REF 5-10106, REF 5-10206 b) 3.5 mm, REF 5-10107 c) 4.0 mm, REF 5-10108, REF 5-102
Hazard: Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
September 18, 2019Medical DevicesFDA
Teleflex Medical recalls Hudson RCI Sheridan LTS: a) 4.0 mm, REF 5-11108 b) 5.0 mm, REF 5-11110 c) 6.0 mm, REF 5-11112 Product Usage: Trac
Hazard: Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
August 7, 2019Medical DevicesFDA
Teleflex Medical recalls Rusch EasyCath Kit
Hazard: Potential for product not to be sterile.
July 24, 2019Medical DevicesFDA
Teleflex Medical recalls RUSCH Urinary Drainage bag, 2000mlm REF 390060 urine collection device
Hazard: The device labels are not UDI compliant. The missing UDI compliance information on the labelling would not affect the function/performance of the device itself. Other label information and warnings are unaffected by this labelling.
July 24, 2019Medical DevicesFDA
Teleflex Medical recalls RUSCH CARE Premium Drain bag, 2000mlm REF 390000 urine collection device
Hazard: The device labels are not UDI compliant. The missing UDI compliance information on the labelling is not likely to affect the function/performance of the device itself. Other label information and warnings are unaffected by this labelling.
July 24, 2019Medical DevicesFDA
Teleflex Medical recalls Teleflex HUDSON RDI NEONATAL ConchaSmart Breathing Circuit with Dual Heated Limb and ConchaSmart Column: a) REF 870-07
Hazard: Reported complaints of cracks being observed, prior to use, on swivel wye adaptors of the Neonatal ConchaSmart Breathing Circuit.
June 26, 2019Medical DevicesFDA
Teleflex Medical recalls Sheridan/HVT: 1) Novaplus 7.0 mm, Product Code V5-10314 2) Novaplus 7.5 mm, Product Code V5-10315 3) Novaplus 8.0 mm,
Hazard: This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
June 26, 2019Medical DevicesFDA
Teleflex Medical recalls Sheridan/CF Novaplus 7.0 mm, Product Code V5-10114 Product Usage: Tracheal tube/airway management
Hazard: This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
June 26, 2019Medical DevicesFDA
Teleflex Medical recalls Hudson RCI Sheridan: 1) HVT 6.0 mm, Product Code 5-10312 2) HVT 6.5 mm, Product Code 5-10313 3) HVT 7.0 mm, Product C
Hazard: This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
June 26, 2019Medical DevicesFDA
Teleflex Medical recalls Hudson RCI Sheridan: 1) Uncuffed 6.0 mm, Product Code 5-10412 2) Uncuffed 6.5 mm, Product Code 5-10413 3) Uncuffed 7.
Hazard: This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
June 26, 2019Medical DevicesFDA
Teleflex Medical recalls Hudson RCI Sheridan: 1) Preformed 6.0 mm, Product Codes: a) 5-22212, b) 5-22312, c) 5-22112, d) 5-22012, 2)
Hazard: This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
June 26, 2019Medical DevicesFDA
Teleflex Medical recalls Hudson RCI Sheridan: 1) EZ-ENDO 6.0 mm, Product Code 5-22512 2) EZ-ENDO 6.5 mm, Product Code 5-22513 3) EZ-ENDO 7.0 m
Hazard: This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
June 26, 2019Medical DevicesFDA
Teleflex Medical recalls Hudson RCI Sheridan: 1) CF 6.0 mm, Product Codes: 5-10112, 5-10212 2) CF 6.5 mm, Product Codes: 5-10113, 5-10213 3)
Hazard: This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
June 26, 2019Medical DevicesFDA
Teleflex Medical recalls Hudson RCI Sheridan LTS, Product Code 5-11112 Product Usage: Tracheal tube/airway management
Hazard: This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.