Tenacore Recalls
2 recalls on record · April 14, 2021 to April 14, 2021 · Medical Devices
April 14, 2021Medical DevicesFDA
Tenacore LLC recalls Alaris Infusion Pump 8100 Bezel (plastic piece only), Part: TIPA-8100-4410
Hazard: Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
April 14, 2021Medical DevicesFDA
Tenacore LLC recalls Alaris Infusion Pump Module, Model 8100, serviced/repaired by Tenacor with an impacted bezel repair part
Hazard: Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.