Terumo Cardiovascular Systems Recalls

223 recalls on record · June 27, 2012 to July 2, 2025 · Medical Devices

Recalls by year

12
13
14
15
16
17
18
19
20
21
22
23
24
25

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

April 23, 2014Medical DevicesFDA
Terumo Cardiovascular Systems Corp recalls Terumo Cardiovascular Procedure Kit- X-Coated Pump and Table Pack with FX25RE Catalog Number: 73431
Hazard: Incorrect customer notification label applied to the Tyvek lid of this specific lot of Terumo¿ Cardiovascular Procedure Kits
March 26, 2014Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Sarns Dual-stage Venous Return Cannulae, 34/46 Fr with 1/2" flare The Sarns Malleable Dual-stage Venous Return Cathet
Hazard: During in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return Cannulae.
March 26, 2014Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Sarns Dual-stage Venous Return Cannulae, 32/40 Fr with 1/2" flare The Sarns Malleable Dual-stage Venous Return Cathet
Hazard: During in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return Cannulae.
March 26, 2014Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Sarns Malleable Dual-stage Venous Return Cannulae, 34/46 Fr with 1/2" connector, 15" (38 cm) long, with X Coating The
Hazard: During in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return Cannulae.
March 26, 2014Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Sarns Malleable Dual-stage Venous Return Cannulae, 34/46 Fr with 1/2" connector, 15" (38 cm) long The Sarns Malleable
Hazard: During in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return Cannulae.
March 26, 2014Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Sarns Malleable Dual-stage Venous Return Cannulae, 32/40 Fr with 1/2" connector, 15" (38 cm) long The Sarns Malleable
Hazard: During in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return Cannulae.
March 26, 2014Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Sarns Malleable Dual-stage Venous Return Cannulae, 28/38 Fr with 1/2" connector, 14.5" (37 cm) long, with X Coating T
Hazard: During in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return Cannulae.
March 26, 2014Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Sarns Malleable Dual-stage Venous Return Cannulae, 28/38 Fr with 1/2" connector, 14.5" (37 cm) long The Sarns Malleab
Hazard: During in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return Cannulae.
March 5, 2014Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls VirtuoSaph¿ Plus Endoscopic Vessel Harvesting System, Part number: VSP550, Sterile***Rx Only*** Product Usage: The V
Hazard: Sterility of medical devices intended for use in surgical procedures may be compromised.
January 8, 2014Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Terumo Cardiovascular Custom Cardiovascular Procedure Kits containing an Overpressure Safety Valve, LN130B, in the Arter
Hazard: Over-Pressure Safety Valves not cleared with an indication for use on the arterial side of the extracorporeal bypass circuit.
November 27, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Vascutek Gelsoft Vascular Graft (peripheral) Labeling reads in part, "***Gelsoft Gelatin Impregnated Knitted Vascular P
Hazard: Terumo Cardiovascular Systems (Terumo CVS) became aware of a customer complaint that a 15mm x 8mm Gelsoft graft was received which the customer believed was a 15mm x 6mm Gelsoft graft unit. The manufacturer confirmed a batch of product code 631508 was incorrectly labeled.
November 20, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Sarns" Venous Return Cannulae, 20 Fr with 1/4" flare, 14.5" (37 cm) long. Indicated for venous drainage during cardi
Hazard: During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications on certain product lots of Sarns" Soft-Flow¿ Aortic Cannulae and Sarns" Venous Return Cannulae.
November 20, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Sarns" Soft-Flow¿ Aortic Cannula without Suture Flange, straight tip, wire-reinforced with luer port, aortic cannula, 8
Hazard: During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications on certain product lots of Sarns" Soft-Flow¿ Aortic Cannulae and Sarns" Venous Return Cannulae.
November 20, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Sarns" Soft-Flow¿ Aortic Cannula without Suture Flange, straight tip, wire-reinforced with luer port, aortic cannula, 7
Hazard: During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications on certain product lots of Sarns" Soft-Flow¿ Aortic Cannulae and Sarns" Venous Return Cannulae.
November 20, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Sarns" Soft-Flow¿ Aortic Cannula without Suture Flange, angled tip, wire-reinforced with luer port, aortic cannula, 8.0
Hazard: During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications on certain product lots of Sarns" Soft-Flow¿ Aortic Cannulae and Sarns" Venous Return Cannulae.
October 2, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Terumo Pall AL6X Arterial Blood Line Filter, labeled in part Terumo Cardiovascular Systems Corporation, packaged as part
Hazard: Medical device used in cardiovascular procedures is defective.
August 21, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls 220/240V AC, Advanced Perfusion System 1 Product Usage: The Terumo¿ Advanced Perfusion System 1 is indicated for use
Hazard: TCVS is taking action to update the Operators Manual with the specific instructions to respond to this condition and will issue a labeling addendum for current users. In the event of a false high pressure alarm condition occurs that displays a "999" pressure reading and results in stopping a pump, this pressure display can be cleared and the pump re-started by disconnecting the pressure transducer from the system.
August 21, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls 100/120V AC, Advanced Perfusion System 1 Product Usage: The Terumo¿ Advanced Perfusion System 1 is indicated for use
Hazard: TCVS is taking action to update the Operators Manual with the specific instructions to respond to this condition and will issue a labeling addendum for current users. In the event of a false high pressure alarm condition occurs that displays a "999" pressure reading and results in stopping a pump, this pressure display can be cleared and the pump re-started by disconnecting the pressure transducer from the system.
August 21, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Terumo¿ Advanced Perfusion System 1 (APS1, System 1) Description: 220/240V AC Product Usage: The Terumo¿ Advanced P
Hazard: Terumo Cardiovascular Systems (TCVS) has become aware that an Advanced Perfusion System 1 (APS1) Centrifugal Pump 'Service Pump' message may occur due to user induced motor movement. In this instance, the APS1 will identify the movement (as it is designed) and display a Service Pump message error. The user might interpret this as a pump which requires service rather than a message displayed due to pump movement while not activated.
August 21, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Terumo¿ Advanced Perfusion System 1 (APS1, System 1) 100/120V AC Product Usage: The Terumo¿ Advanced Perfusion Syst
Hazard: Terumo Cardiovascular Systems (TCVS) has become aware that an Advanced Perfusion System 1 (APS1) Centrifugal Pump 'Service Pump' message may occur due to user induced motor movement. In this instance, the APS1 will identify the movement (as it is designed) and display a Service Pump message error. The user might interpret this as a pump which requires service rather than a message displayed due to pump movement while not activated.
August 21, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Terumo¿ Advanced Perfusion System 1: 220/240V AC, Advanced Perfusion System 1 The Terumo¿ Advanced Perfusion System
Hazard: The operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the Central Control Monitor at the same time an alarm condition occurs with configuration for a pump coast response, the labeling in the operators manual does not fully describe how the pump will react.
August 21, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Terumo¿ Advanced Perfusion System 1: 100/120V AC, Advanced Perfusion System 1 The Terumo¿ Advanced Perfusion System 1
Hazard: The operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the Central Control Monitor at the same time an alarm condition occurs with configuration for a pump coast response, the labeling in the operators manual does not fully describe how the pump will react.
July 24, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls CDI Blood Parameter Monitoring System 500 with Venous blood parameter module. Intended for use during cardiopulmonary
Hazard: Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains inaccuracies associated with a software algorithm, and a software update has been developed to address this issue and a premarket notification (K130425) has been submitted to the agency. Once clearance is received for this premarket notification from the US FDA, Terumo CVS will implement the
July 24, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls CDI Blood Parameter Monitoring System 500 with Arterial blood parameter module. Intended for use during cardiopulmona
Hazard: Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains inaccuracies associated with a software algorithm, and a software update has been developed to address this issue and a premarket notification (K130425) has been submitted to the agency. Once clearance is received for this premarket notification from the US FDA, Terumo CVS will implement the
July 24, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls CDI Blood Parameter Monitoring System 500 with Arterial and Venous blood parameter modules. Intended for use during car
Hazard: Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations.
July 24, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls CDI Blood Parameter Monitoring System 500 with Arterial blood parameter modules and Hematocrit/Saturation probe. Int
Hazard: Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains inaccuracies associated with a software algorithm, and a software update has been developed to address this issue and a premarket notification (K130425) has been submitted to the agency. Once clearance is received for this premarket notification from the US FDA, Terumo CVS will implement the
July 24, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls CDI Blood Parameter Monitoring System 500 with Arterial and Venous blood parameter modules and Hematocrit/Saturation pro
Hazard: Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains inaccuracies associated with a software algorithm, and a software update has been developed to address this issue and a premarket notification (K130425) has been submitted to the agency. Once clearance is received for this premarket notification from the US FDA, Terumo CVS will implement the
July 3, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Sarns " TCM II Temperature Control Module The Sarns" TCM II Temperature Control Module is a source of temperature-con
Hazard: Internal testing found certain Sarns" TCM II systems were released with incorrectly assembled heatsink assemblies which could cause the units to fail.
February 20, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring System. Monitors blood/oxygen saturation, hematocrit and hemoglob
Hazard: Inaccurate readings. Software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations. The accuracy on any cuvette size did not meet the specifications.
January 23, 2013Medical DevicesFDA
Terumo Cardiovascular Systems Corporation recalls Sarns high-flow aortic arch cannula, 6.5 mm (20 Fr) OD with 3/8in connector, 11in (28 cm) long, with X coating surface
Hazard: Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part.