261 recalls on record · August 1, 2012 to January 18, 2023 · Medical Devices
Recalls by year
12
13
14
15
16
17
18
23
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
May 28, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls Anspach Irrigation Clip 40 Product Usage: Irrigation clips provide a method of supplying controlled, cooling irrigat
Hazard: The Irrigation Clips were assembled with the incorrect (smaller) clips.
May 21, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls ANSPACH Cutting bur, 3 mm Diamond Ball, Standard Length. REF L-3SD Use with LONG Attachments. Used for cutting and shapi
Hazard: Anspach product code L-3SD contained product code S-3SD which is shorter in length than the labeled product.
February 5, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls Anspach REF: TURQ-434-4 Twist Drill for 1.5mm x 4mm screw. Rx only Single use Sterile bone cutting burs.
Hazard: Cutter was longer than was stated on the label
February 5, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls AnspacSingle use Sterile bone cutting burs.h REF: S-1504TD Twist Drill for 1.5mm x 4mm screw. Rx only Single use Steri
Hazard: Cutter was longer than was stated on the label
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls Anspach MIA 16 (16cm Minimally Invasive Attachment), Rx Only MIA16 attachments are intended for use with Motor System
Hazard: Product did not reach expected sterility assurance level
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA7-2SD - 2MM Diamond Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping delicat
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA7-2SB - 2MM Fluted Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping delicate
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA7-1SD - 1MM Diamond Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping delicat
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA7-1SB - 1MM Fluted Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping delicate
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA7-15SD - 1.5MM Diamond Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping deli
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA7-15SB - 1.5MM Fluted Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping delic
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA7-08SD - 0.8MM Diamond Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping deli
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA7-07SD - 0.7MM Diamond Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping deli
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA7-06SD - 0.6MM Diamond Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping deli
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA7-05SD - 0.5MM Diamond Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping deli
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA5-2SD - 2MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicat
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA5-2SB - 2MM Fluted Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA5-1SD - 1MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicat
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA5-1SB - 1MM Fluted Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA5-15SD - 1.5MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping deli
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA5-15SB - 1.5MM Fluted Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delic
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA5-08SD - 0.8MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping deli
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA5-07SD - 0.7MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping deli
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA5-06SD - 0.6MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping deli
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls MCA5-05SD - 0.5 MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping del
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls Micro Curved Attachment (MCA) Cutting Burr, DFU 18-0001. All lots are included in recall. Part Number MCA Intended f
Hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls Anspach Carbide Cutting Burrs & Drums: 2MM Carbide Fluted Ball, Extended part # S-2B-C and 7MM Carbide Fluted Ball part
Hazard: Some carbide cutting burr fractured during use.
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls Anspach Carbide Cutting Burrs & Drums: 1.5MM x 7.5MM Cylindrical Carbide Drum part # S-15DRMC Anspach cutting burrs
Hazard: Some carbide cutting burr fractured during use.
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls Anspach Carbide Cutting Burrs: 1.5MMx7.5MM Cylindrical Carbide Drum part # S-15DRMC-L, 1MM Carbide Fluted ball, Extende
Hazard: Some carbide cutting burr fractured during use.
January 22, 2014Medical DevicesFDA
The Anspach Effort, Inc. recalls Anspach Carbide Cutting Burrs: 2MM Carbide Fluted Ball, Standard, 8 Flutes - part #QD8-2SB-C-8F, 4MM Carbide Fluted Bal
Hazard: Some carbide cutting burr fractured during use.