The Metrix Recalls
33 recalls on record · August 3, 2016 to November 20, 2024 · Medical Devices
Recalls by year
16
17
19
24
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
August 3, 2016Medical DevicesFDA
The Metrix Company recalls AUTOCOMP6 High Speed Compounder
Hazard: An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.
August 3, 2016Medical DevicesFDA
The Metrix Company recalls AUTOCOMP6 XP High Speed Compounder REF 58810
Hazard: An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.
August 3, 2016Medical DevicesFDA
The Metrix Company recalls AUTOCOMP6 XPS High Speed Compounder REF 58810
Hazard: An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.