July 15, 2026Medical DevicesFDA
Think Surgical, Inc. recalls TMINI Femur Tracking Array REF 109003 UDI-DI code: M9461090030 TMINI Tibia Tracking Array REF 109004 UDI-DI code: M
Hazard: Due to manufacturing/design issue where the knob on the tracking array may loosen causing potential misalignment of the array.
March 27, 2024Medical DevicesFDA
Think Surgical, Inc. recalls Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic S
Hazard: There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling.
August 15, 2018Medical DevicesFDA
Think Surgical, Inc. recalls Think Surgical Cutter, Flat, ¿6.2mm x 144mm. For use only with the TCAT Computer-Assisted Tool. Part Number 106428
Hazard: There is a potential for the cutter head to detach from the cutter shaft.