Home / Brands / Tosoh Bioscience
Tosoh Bioscience Recalls
74 recalls on record · April 17, 2013 to July 22, 2026 · Medical Devices
Recalls by year
13
15
16
17
18
19
20
21
26
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
July 22, 2026Medical DevicesFDA
Tosoh Bioscience Inc recalls ST AIA-Pack Testosterone, IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testosterone in human serum o
Hazard: The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated or falsely low total testosterone results. Such erroneous test results may lead to additional testing/diagnostic interventions and in limited cases, may lead to a delay in diagnosis and/or initiation of unnecessary or incorrect clinical treatment measures.
March 17, 2021Medical DevicesFDA
Tosoh Bioscience Inc recalls Thyroglobulin Antibody (TgAB) Calibrator - Product Usage: intended for IN VITRO DIAGNOSTIC USE ONLY for the quantitativ
Hazard: Some combinations of Calibrator lots and AlA PACKS resulted in Calibrator #6 not giving a value, making the calibration unacceptable.
September 23, 2020Medical DevicesFDA
Tosoh Bioscience Inc recalls Tosoh Bioscience AIA-PACK BHCG Calibration Verification Test Set-In Vitro Diagnostic for BHCG-pregnancy Part number: 02
Hazard: Decreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up to 20%.
September 23, 2020Medical DevicesFDA
Tosoh Bioscience Inc recalls Tosoh Bioscience AIA- PACK HCG Calibrator Set-In Vitro Diagnostic for BHCG Part number 020361 The AIA-Pack BHCG Cali
Hazard: Decreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up to 20%.
July 1, 2020Medical DevicesFDA
Tosoh Bioscience Inc recalls AIA-900 Analyzer, Part no. 022930, UDI 04560189283992
Hazard: The Substrate Monitoring System in the AIA-900 Analyzer did not detect an empty substrate bottle. The issue may lead to erroneous immunoassay test results.
March 11, 2020Medical DevicesFDA
Tosoh Bioscience Inc recalls AIA-360 Automated Immunoassay Analyzer, Product Code 019945
Hazard: A display screen software issue on the analyzer causes the display to freeze when display screen is touched at the same time as a command from the instrument firmware, causing the instrument to stop. As a result, the run is aborted and the results are not retrievable.
January 2, 2019Medical DevicesFDA
Tosoh Bioscience Inc recalls ST AIA-PACK ACTH (ACTH: Adrenocorticotropic Hormone) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative m
Hazard: FDA Class II
January 2, 2019Medical DevicesFDA
Tosoh Bioscience Inc recalls ST AIA-PACK Homocysteine (HCY) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of homocyst
Hazard: FDA Class II
January 2, 2019Medical DevicesFDA
Tosoh Bioscience Inc recalls ST AIA-PACK Testosterone (TES) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of testoste
Hazard: FDA Class II
January 2, 2019Medical DevicesFDA
Tosoh Bioscience Inc recalls ST AIA-PACK FT3 (FT3: Free Triiodothyronine) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measureme
Hazard: FDA Class II
January 2, 2019Medical DevicesFDA
Tosoh Bioscience Inc recalls AIA-PACK Folate (FOL) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of folate in human s
Hazard: FDA Class II
January 2, 2019Medical DevicesFDA
Tosoh Bioscience Inc recalls AIA-PACK B12 (Vitamin B12) is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of vitamin B12
Hazard: Potential for fluorescein interference to cause inaccurately elevated results for certain Tosoh AIA assays
December 26, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls AIA-360 Analyzer is capable of performing two methods of immunoassay: an immunoenzymetric (IEMA) or sandwich immunoassay
Hazard: The firm has become aware of a potential issue with the AIA-360 analyzer that could prevent the table home sensor from detecting the table home position. If the malfunction with the table home sensor occurs on the AIA-360, the sensor board may need to be replaced. If the home sensor fails to detect the home position, the AIA-360 analyzer will cease to function and will be unavailable for use. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received three (3) complaints related to this issue with no serious injuries reported.
December 19, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls AIA-900 Automated Immunoassay Analyzer performs three methods of immunoassay: an immunoenzymetric (IEMA) or sandwich im
Hazard: The firm has become aware of a potential issue that may affect the mixing motor drive harness and mixing home detection sensor relay harness on your AIA-900 Analyzer. It is possible that the harnesses could rub against the Plarail chain during operation and result in premature wear. If observed, this issue will cause the following system error to be reported D.lane Mix home overrun error [4093] . If observed, this issue will cause the AIA-900 Analyzer not to function and the D.lane Mix home overrun error [4093] will be reported by the system. Replacement of the motor drive harness and mixing home detection sensor relay harness will mitigate the potential for premature wear and effectively prevent this issue from causing the system error. If the sensor relay harness fails, the AIA-900 analyzer will cease to function and will be unavailable for use. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh is aware of two complaints associated with the unexpected wearing of the harness. There have been no user or patient injuries reported that are related to this device malfunction.
December 19, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls Hemoglobin A1c Control, Part no. 220232 for use as quality control material to monitor the precision of laboratory testi
Hazard: Tosoh Bioscience has become aware of potential degradation of HbA1c control lot # 7055 before the 60-days product stability claim. If degradation of these controls occurs, P00 peaks may be seen and the % HbA1c control values may drift higher than the assigned range. This may result in delays when performing routine instrument quality control. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received two (2) complaints related to this issue with no serious injuries reported.
December 19, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls AIA-900Analyzer, Product code 022930, 022930R The AIA-2000 analyzer and AIA-900 analyzer are capable of performing th
Hazard: Tosoh Bioscience has become aware of a potential issue with the AIA-900 and AIA-2000 analyzers that could result in slippage of the incubator unit contained within the system. If the tension of the timing belt is not adjusted correctly, it is possible for a misalignment that will result in a mechanical error. Tosoh has taken the required actions to inspect the tension of the timing belt during preventative maintenance and correct the instruments where this issue has been observed. If incubator slippage occurs, the unit will cease to function, and a system error will be reported. If this occurs, there may be a delay in testing and reporting patient test results. There have been no reports of injury related to this issue.
December 19, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls AIA-2000 Analyzer, Product code 022100, 022100R, 022101, 022101R The AIA-2000 analyzer and AIA-900 analyzer are capa
Hazard: Tosoh Bioscience has become aware of a potential issue with the AIA-900 and AIA-2000 analyzers that could result in slippage of the incubator unit contained within the system. If the tension of the timing belt is not adjusted correctly, it is possible for a misalignment that will result in a mechanical error. Tosoh has taken the required actions to inspect the tension of the timing belt during preventative maintenance and correct the instruments where this issue has been observed. If incubator slippage occurs, the unit will cease to function, and a system error will be reported. If this occurs, there may be a delay in testing and reporting patient test results. There have been no reports of injury related to this issue.
December 12, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls AIA-360 Analyzer, product code 019945/019945R The Tosoh AIA-360 Automated Immunoassay Analyzer is for use in vitro d
Hazard: The firm has implemented new product labeling changes for the Tosoh AIA-360 Analyzer. These labeling changes affect the Tosoh Quick Reference Guide, Training Manual and Training DVD previously provided by Tosoh Bioscience. These changes are being made to align the Tosoh Quick Reference Guide and associated training materials to the Operators Manual and provide clarification to the requirement that the maintenance of the B/F probe is to be performed by Tosoh a field service representative, the requirement for use of 70% Ethanol during daily shutdown procedures and the requirement for use of CAP Class 1 reagent grade water for dilution of the concentrated wash and diluent solutions and reconstitution of lyophilized reagents. Failure to follow these requirements may negatively affect system performance, lead to corrosion of instrument components over time, or result in bacterial contamination, respectively.
December 5, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls AIA-900 Automated Immunoassay Analyzer, product code 022930 /022930R Product Usage: The AIA-900 is capable of perfor
Hazard: Tosoh Bioscience has become aware of a potential issue with the axis base of the turntable rotation drive motor on the AIA-900 Analyzer. If the tension of the timing belt between the turntable rotation pulley and drive motor is not adjusted correctly, it is possible to increase the lateral load on the motor axis resulting in the inclination and damage of the motor axis. Since January 12, 2017, Tosoh has been inspecting the tension of the timing belt during scheduled preventative maintenance visit and correcting the analyzers where this issue has been observed. If the drive motor fails, the AIA-900 analyzer will cease to function and will be unavailable for use. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received five (5) complaints related to this issue with no serious injuries reported.
November 28, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls Tosoh AIA-900 immunoassay Analyzer
Hazard: A software issue may cause the liquid volume detection entries for 2nd and 3rd reagent bottles to be overwritten with the data from the 1st reagent bottle. Tosoh has released software version 2.51 (or later) to fix the problem. This failure is mitigated by the hardware design. The consequence would be a slight delay in testing patient sample or reporting patient result.
November 28, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls AIA-2000 Analyzer, Model Nos. AIA-2000LA, AIA-2000ST - Product Usage: The AIA-2000 is capable of performing three metho
Hazard: There is a possibility that the liquid suction pump on some AIA-2000 analyzers may not perform as intended and cause an insufficient volume of wash solution to be removed during the washing step. Incomplete washing of the sample may result in falsely elevated test results for the sandwich assays and falsely decreased test results for the competitive binding assays. Insufficient washing is detectable through review of the calibration data and QC trending. For example, if incomplete washing occurs, the reported values for the zero- calibration rate will be above the acceptance range. The calibration value may shift and the instrument will not calibrate successfully. If this issue occurs, there may be a delay in testing and reporting patient test results. There have been no user or patient injuries reported related to this issue.
November 21, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls AIA-900 Analyzer, Product code 022930, 022930R
Hazard: Tosoh Bioscience has become aware of a potential issue with the pickup arm and pickup arm motor on the AIA-900 Analyzer. If the pickup arm or pickup arm motor malfunctions, the analyzer will cease to function and result in an inability to test patient samples. This may cause a delay in testing and reporting patient results. Tosoh has received fifty-one (51) complaints related to this issue with no serious injuries reported.
October 3, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls G8 Automated HPLC Analyzer - 723G8 Product Usage - In vitro diagnostic use for the quantitative measurement of % hemo
Hazard: Analyzers were distributed with software which lacks a 510K
August 22, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls Tosoh Bioscience AIA Analyte Application Manual (AAM) CD Part Number: 997027 CD contains analyte application manuals
Hazard: Distribution of unapproved product labeling
August 22, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls G8 Automated HPLC Analyzer: HLC-723G8-ST, 021560; and HLC-723G8-LA, 021674 Product Usage: The Tosoh Automated Glycoh
Hazard: HbAE is known to interfere with the HbA1c assay on the current version of software, Ver. 5.23. Customers should exercise caution when reviewing chromatograms and ensure that Flag 43 is enabled on their device to avoid reporting invalid test results in the presence of HbAE.
June 13, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls ST-AIA PACK Cystatin C; Part Number: 025217 Assay, Kidney Marker
Hazard: Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.
June 13, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls ST-AIA PACK BMG; Part Number:025259 Assay, Tumor Marker
Hazard: Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.
June 13, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls ST-AIA PACK IGE II; Part Number: 025295 Assay, Allergy
Hazard: Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.
June 13, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls ST-AIA PACK Fer; Part Number: 025253 Assay, Anemia
Hazard: Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.
June 13, 2018Medical DevicesFDA
Tosoh Bioscience Inc recalls ST-AIA PACK HGH; Part Number: 025266 Assay, Metabolic
Hazard: Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.