Visaris DOO Recalls
3 recalls on record · May 23, 2018 to May 23, 2018 · Medical Devices
May 23, 2018Medical DevicesFDA
Visaris DOO recalls Visaris Vision/Auto, ddRAura OTC APS, Diagnostic X-Ray System, Version 2.4 The Vision systems are intended for use in
Hazard: A failure of the tube arm suspension is possible while the tube arm height is being adjusted. During tube height adjustment there is a small probability of failure of both the primary and redundant holding mechanisms, causing uncontrollable descent of the tube arm and a potential risk of injury to the patient.
May 23, 2018Medical DevicesFDA
Visaris DOO recalls Visaris Vision/Auto, ddRAura OTC APS, Diagnostic X-Ray System, Version 2.3 The Vision systems are intended for use in a
Hazard: A failure of the tube arm suspension is possible while the tube arm height is being adjusted. During tube height adjustment there is a small probability of failure of both the primary and redundant holding mechanisms, causing uncontrollable descent of the tube arm and a potential risk of injury to the patient.
May 23, 2018Medical DevicesFDA
Visaris DOO recalls Visaris Vision/Auto, ddRAura OTC APS, Diagnostic X-Ray System, Version 2.2 The Vision systems are intended for use in
Hazard: A failure of the tube arm suspension is possible while the tube arm height is being adjusted. During tube height adjustment there is a small probability of failure of both the primary and redundant holding mechanisms, causing uncontrollable descent of the tube arm and a potential risk of injury to the patient.