Home / Brands / Vortran Medical Technology 1
Vortran Medical Technology 1 Recalls
2 recalls on record · May 24, 2017 to May 24, 2017 · Medical Devices
May 24, 2017Medical DevicesFDA
Vortran Medical Technology 1, Inc recalls VORTRAN Automatic Resuscitator, VARPlus with Extension Kit; Model PCE-5012; Qty: 10 each. Anesthesiology: This devic
Hazard: Vortran received a customer complaint that the spin-nut-DSS connector was assembled in the reverse direction on the oxygen tubing; making the connector unable to be screwed on due to the wrong thread direction.
May 24, 2017Medical DevicesFDA
Vortran Medical Technology 1, Inc recalls VORTRAN Automatic Resuscitator, VARPlus; Model PTE-5002; Qty: 10 each Anesthesiology: This device is to be used by p
Hazard: Vortran received a customer complaint that the spin-nut-DSS connector was assembled in the reverse direction on the oxygen tubing; making the connector unable to be screwed on due to the wrong thread direction.