Vyaire Medical Recalls
63 recalls on record · September 6, 2017 to May 21, 2025 · Medical Devices
Recalls by year
17
18
19
20
22
23
24
25
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
May 21, 2025Medical DevicesFDA
Vyaire Medical recalls Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH265 Software Version: N/
Hazard: Adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.
May 21, 2025Medical DevicesFDA
Vyaire Medical recalls Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH165 Software Version: N/
Hazard: Adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.
May 15, 2024Medical DevicesFDA
Vyaire Medical, Inc. recalls Vyaire Medical GmbH, Vyaire Twin Tube sample line 240 cm, Model Number V-707327; The TwinTube sample line 240cm is used
Hazard: Potential of the nozzle separating during patient use.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K8004
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, Variable Volume Oxygen Reservoir Tubing, Adult Mask, REF 2K8017
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C2
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, Expiratory Filter, REF 2K8005F
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K8005
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Adult Mask, CO2 Detector, REF 2K8004C2
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Expiratory Filter, REF 2K8004F
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, without Mask, REF 2K8001
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, without Mask, REF 2K8000,
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
December 13, 2023Medical DevicesFDA
Vyaire Medical recalls vyaire Flexible Patient Circuit, 3100A, REF 29028-004, intended to be used with the 3100A High Frequency Oscillatory Ven
Hazard: Two patient circuits, 29028-003 and 29028-004, for the 3100 High Frequency Oscillatory ventilator that are only approved for use outside the U.S. were distributed in the U.S.
December 13, 2023Medical DevicesFDA
Vyaire Medical recalls vyaire Flexible Patient Circuit, 3100A, REF 29028-003, intended to be used with the 3100A High Frequency Oscillatory Ven
Hazard: Two patient circuits, 29028-003 and 29028-004, for the 3100 High Frequency Oscillatory ventilator that are only approved for use outside the U.S. were distributed in the U.S.
June 21, 2023Medical DevicesFDA
Vyaire Medical recalls bellavista 1000e Ventilator, Catalogue Number 301.100.130; Continuous Use Ventilator
Hazard: Vyaire Medical identified two patient safety risks during the use of the bellavista 1000 and bellavista 1000e: 1. Under certain conditions of use, the touchscreen may become unresponsive and the device application stops responding (APP Hang). When this occurs, if the user continues to interact with the touchscreen, a message will appear that states DeviceSoftware.Application is not responding or a decommission screen will appear on the user interface. The ventilator will issue both an audible and visual alarm. Ventilation continues without interruption with the settings applied prior to the APP Hang error. 2. The potential for a use error has been identified if the operator applies the proposed settings without confirming the settings are suitable for the patient.
June 21, 2023Medical DevicesFDA
Vyaire Medical recalls bellavista 1000 Ventilator, Catalogue Number 301.100.030; Continuous Use Ventilator
Hazard: Vyaire Medical identified two patient safety risks during the use of the bellavista 1000 and bellavista 1000e: 1. Under certain conditions of use, the touchscreen may become unresponsive and the device application stops responding (APP Hang). When this occurs, if the user continues to interact with the touchscreen, a message will appear that states DeviceSoftware.Application is not responding or a decommission screen will appear on the user interface. The ventilator will issue both an audible and visual alarm. Ventilation continues without interruption with the settings applied prior to the APP Hang error. 2. The potential for a use error has been identified if the operator applies the proposed settings without confirming the settings are suitable for the patient.
February 23, 2022Medical DevicesFDA
Vyaire Medical recalls bellavista 1000 ventilator, Catalog #301.100.030 and bellavista 1000e ventilator, 301.100.130. The bellavista 1000/10
Hazard: Potential cessation in ventilation can occur under specific conditions.
March 11, 2020Medical DevicesFDA
Vyaire Medical recalls bellavista 1000 ventilator, Catalog number 301.100.030. bellavista is an electronically controlled pneumatic ventilatio
Hazard: The G6 bellavista 1000 US ventilators may experience intermittent failures: Lack of acoustic high priority alarm, presence of a 'no alarm' condition, or presence of non-responsive touch screen.
July 3, 2019Medical DevicesFDA
Vyaire Medical recalls enFlow Disposable Cartridge with IV Extension Set, used for intravenous warming therapy with fluid and blood solutions.
Hazard: Testing has demonstrated aluminum elution from the enFlow Disposable Cartridge during fluid warming.
July 3, 2019Medical DevicesFDA
Vyaire Medical recalls enFlow Disposable Cartridge, used for intravenous warming therapy with fluid and blood solutions.
Hazard: Testing has demonstrated aluminum elution from the enFlow Disposable Cartridge during fluid warming.
August 29, 2018Medical DevicesFDA
Vyaire Medical recalls The Tri-Flo Subglottic Suction System
Hazard: Vyaire Medical has discovered potential patient safety risk with associated with the Tri-Flow Subglottic Suction System. The distal soft tip of the catheter has been identified as having the potential to break off during patient use and result in aspiration of a solid foreign body.
August 15, 2018Medical DevicesFDA
Vyaire Medical recalls AirLife Resuscitation Device, Pediatric labeled as the following: a. Pediatric, with mask, 28" large bore tubing, p
Hazard: An error in manufacturing of the AirLife Resuscitation Devices that has the potential to obstruct airflow of the supplemental oxygen delivery during use.
August 15, 2018Medical DevicesFDA
Vyaire Medical recalls AirLife Resuscitation Device, Adult labeled as the following: a. Adult, with mask, oxygen reservoir bag, Part Numbe
Hazard: An error in manufacturing of the AirLife Resuscitation Devices that has the potential to obstruct airflow of the supplemental oxygen delivery during use.
May 16, 2018Medical DevicesFDA
Vyaire Medical recalls AirLife Resuscitation, Infant, with mask, oxygen reservoir bag, pressure-relief valve, PEEP valve Product Usage: The
Hazard: The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
May 16, 2018Medical DevicesFDA
Vyaire Medical recalls AirLife Resuscitation, Infant, with mask, 40inch oxygen reservoir tubing, pressure-relief valve, manometer, PEEP valve
Hazard: The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
May 16, 2018Medical DevicesFDA
Vyaire Medical recalls AirLife Resuscitation, Replacement kit for outer shell, Broselow Product Usage: The AirLife Resuscitation devices ar
Hazard: The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
May 16, 2018Medical DevicesFDA
Vyaire Medical recalls AirLife Resuscitation, Broselow Hinkle Kit Product Usage: The AirLife Resuscitation devices are manual resuscitation
Hazard: The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
May 16, 2018Medical DevicesFDA
Vyaire Medical recalls AirLife Resuscitation, Broselow Hinkle Kit Product Usage: The AirLife Resuscitation devices are manual resuscitation
Hazard: The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
May 16, 2018Medical DevicesFDA
Vyaire Medical recalls AirLife Resuscitation, Pediatric, with mask, 40inch oxygen reservoir tubing, pressure-relief valve, Broselow Product
Hazard: The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
May 16, 2018Medical DevicesFDA
Vyaire Medical recalls AirLife Resuscitation, Pediatric, with mask, oxygen reservoir bag, manometer, pressure-relief valve, PEEP valve Produ
Hazard: The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.