Xstrahl Recalls
3 recalls on record · November 3, 2021 to July 27, 2022 · Medical Devices
Recalls by year
21
22
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
July 27, 2022Medical DevicesFDA
Xstrahl Limited recalls X80 RADiant Photoelectric Therapy System
Hazard: There is a potential compatibility issue with the systems and replacement treatment applicators.
December 22, 2021Medical DevicesFDA
Xstrahl Limited recalls Treatment applicator for Gulmay Medical 150 or Xstrahl 150 superficial X-ray therapy system
Hazard: Base plate may detach from the main body of the treatment applicator.
November 3, 2021Medical DevicesFDA
Xstrahl Limited recalls Concerto User Interface Software provided with the following systems: (1) Xstrahl 100 Electronic Brachytherapy - 100k
Hazard: If a saved treatment plan with 2 opposing beams is edited prior to approval, then Beam 2 is not updated with the changed parameters upon selecting save, resulting in error messages during the treatment and possible mis-treatment.