Home / Brands / Zap Surgical Systems
Zap Surgical Systems Recalls
2 recalls on record · September 21, 2022 to October 8, 2025 · Medical Devices
Recalls by year
22
25
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
October 8, 2025Medical DevicesFDA
Zap Surgical Systems recalls ZAP-X Radiosurgery System, Model: 300150
Hazard: If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.
September 21, 2022Medical DevicesFDA
Zap Surgical Systems recalls ZAP-X Radiosurgery System
Hazard: Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.