Zest Anchors Recalls
2 recalls on record · February 12, 2014 to July 1, 2020 · Medical Devices
Recalls by year
14
20
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
July 1, 2020Medical DevicesFDA
Zest Anchors LLC recalls LOCATOR OVERDENTURE IMPLANT SYSTEM STERILE R Rx Only QTY 1 .
Hazard: The abutment provided with the Implant System may be out of specification, which could lead to a small gap occurring between the implant and abutment on assembly in the patient.
February 12, 2014Medical DevicesFDA
Zest Anchors LLC recalls LOCATOR¿ Implant Attachment System. The LOCATOR¿ Implant Attachment System is a universal hinge, resilient attachment
Hazard: Zest Anchors, LLC is voluntarily recalling specific lots of LOCATOR¿ Implant Attachment Systems because it is currently provided with a label that indicates that the final torque applied to the abutment should be 30 N-cm. It was recently noted by Zest that in order to provide an optimal performance by increasing the factor of safety between the failure load of the abutment screw and the recommend