August 9, 2023Medical DevicesFDA
Scican Ltd. recalls STATCLAVE G4 Chamber Autoclave -Tabletop steam sterilizer intended for use by a health care provider to sterilize medica
Hazard: The STATCLAVE G4 is a dynamic air removal (prevacuum) tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The STATCLAVE G4 has not been designed to sterilize liquid loads, bio medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and / or damage to the autoclave.
August 9, 2023Medical DevicesFDA
Scican Ltd. recalls STATCLAVE G4 Chamber Autoclave -Tabletop steam sterilizer intended for use by a health care provider to sterilize medica
Hazard: The STATCLAVE G4 is a dynamic air removal (prevacuum) tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The STATCLAVE G4 has not been designed to sterilize liquid loads, bio medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and / or damage to the autoclave.
August 9, 2023Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 M20. X-ray Image guided Therapy System.
Hazard: The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.
August 9, 2023Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10 and FD20. X-ray Imaging System.
Hazard: The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.
August 9, 2023Medical DevicesFDA
Covidien, LLC recalls Covidien Cytosponge Cell Collection Device, REF: CYTO-201; and Covidien Cytosponge Cell Collection Kit, REF: CYTO-KIT-US
Hazard: Cell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which could lead to device fragments in patient, obstruction, airway obstruction, secondary intervention, secondary intervention (with the primary procedure), supraglottic airway obstruction, and aspiration.
August 9, 2023Medical DevicesFDA
Abiomed, Inc. recalls Impella RP Flex with SmartAssist intracardiac microaxial blood pump, Product Number 1000323
Hazard: A higher than expected rate of thrombus formation or deposition has been observed. The risk is for thrombus formations or deposits on indwelling central venous lines or cannulas may break free and enter into the Impella RP Flex, resulting in reduced flow, loss of support, or hemolysis
August 9, 2023Medical DevicesFDA
Hamamatsu Corporation recalls NanoZoomer S360MD Slide scanner system-automated system for creating, viewing, and managing digital slides. The NanoZoom
Hazard: If the wrong barcode information is recorded in the NDPi file, a pathologist may refer to the whole slide image (WSI) of a different patient.
August 9, 2023Medical DevicesFDA
GE OEC Medical Systems, Inc recalls Field Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA; P/N 5485304 SERVICE PART, IMAGE INTENSIFIER, 9 INCH, TOSHIBA; P/N
Hazard: Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
August 9, 2023Medical DevicesFDA
GE OEC Medical Systems, Inc recalls OEC Elite Systems with 9-inch Image Intensifier
Hazard: Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
August 9, 2023Medical DevicesFDA
GE OEC Medical Systems, Inc recalls OEC 9900 Systems with 9-inch Image Intensifier
Hazard: Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
August 9, 2023Medical DevicesFDA
GE OEC Medical Systems, Inc recalls OEC 9800 Systems with 9-inch Image Intensifier
Hazard: Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
August 9, 2023Medical DevicesFDA
GE OEC Medical Systems, Inc recalls OEC Flexiview 8800 Systems with 9-inch Image Intensifier
Hazard: Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
August 9, 2023Medical DevicesFDA
Maquet Cardiovascular, LLC recalls ACROBAT-i Vacuum Stabilizer System, Part Number C-OM-10000. Component of a chest stabilization system for cardiac and ch
Hazard: The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
August 9, 2023Medical DevicesFDA
Maquet Cardiovascular, LLC recalls ACROBAT V Vacuum Off-Pump System, Part Number C-OM-9100S. Component of a chest stabilization system for cardiac and ches
Hazard: The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
August 9, 2023Medical DevicesFDA
Maquet Cardiovascular, LLC recalls ACROBAT SUV Vacuum Off-Pump System, Part Number C-OM-9000S. Component of a chest stabilization system for cardiac and c
Hazard: The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
August 9, 2023Medical DevicesFDA
Maquet Cardiovascular, LLC recalls ACCESSRAIL Platform (Standard Blade) Part Number SB-1000, and (Deep Blades) Part Number C-DB-1000. Component of a chest
Hazard: The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
August 9, 2023Medical DevicesFDA
SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION recalls Over-Head Ceiling Tube Support, CH-200, CH-200M, Used with RadSpeed and RadSpeed PRO, Sonial Vision safire, Versa and G4
Hazard: Eight fixing bolts at the base of the X-ray ceiling tube column assembly may be loose or missing so service is required to install a cable, bracket, and label behind the bracket, indicating when service is required, to eliminate the risk of the device falling.
August 9, 2023Medical DevicesFDA
Olympus Corporation of the Americas recalls Always-On Tip Tracked Instruments (SPiN Drive instruments) SPiN Xtend 2.0mm OD, 21ga Intended for use in clinical int
Hazard: Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
August 9, 2023Medical DevicesFDA
Olympus Corporation of the Americas recalls Always-On Tip Tracked Instruments (SPiN Drive instruments) 22ga SPiN Flex Needle, 12mm L, 1.8mm OD-Intended for use in c
Hazard: Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
August 9, 2023Medical DevicesFDA
Olympus Corporation of the Americas recalls Always-On Tip Tracked Instruments (SPiN Drive instruments) Triple Needle Brush, 12mm L, 1.8mm OD-Intended for use in cli
Hazard: Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
August 9, 2023Medical DevicesFDA
Olympus Corporation of the Americas recalls Always-On Tip Tracked Instruments (SPiN Drive instruments) 21ga ANSO Cytology Needle, 12mm L, 1.8mm OD-Intended for use
Hazard: Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
August 9, 2023Medical DevicesFDA
Olympus Corporation of the Americas recalls Always-On Tip Tracked Instruments (SPiN Drive instruments) 19ga ANSO Histology Needle, 12mm L, 1.8mm OD-Intended for use
Hazard: Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
August 9, 2023Medical DevicesFDA
Olympus Corporation of the Americas recalls Always-On Tip Tracked Instruments (SPiN Drive instruments) Forceps, 1.8mm OD, Serrated Cup-Intended for use in clinical
Hazard: Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
August 9, 2023Medical DevicesFDA
Olympus Corporation of the Americas recalls Always-On Tip Tracked Instruments (SPiN Drive instruments) Forceps, 1.8 mm OD, Oval Cup-Intended for use in clinical int
Hazard: Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
August 9, 2023Medical DevicesFDA
Olympus Corporation of the Americas recalls Always-On Tip Tracked Instruments (SPiN Drive¿ instruments) Brush, 15mm L, 1.8mm OD Intended for use in clinical interv
Hazard: Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
August 9, 2023Medical DevicesFDA
Philips North America recalls Incisive CT Plus-Computed Tomography X-Ray System Model: 728149
Hazard: Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down
August 9, 2023Medical DevicesFDA
Philips North America recalls Incisive CT Power (China)-Computed Tomography X-Ray System Model: 728148
Hazard: Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down
August 9, 2023Medical DevicesFDA
Philips North America recalls Incisive CT for Brazil SKD-Computed Tomography X-Ray System Model: 728146
Hazard: Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down
August 9, 2023Medical DevicesFDA
Philips North America recalls Incisive CT-Computed Tomography X-Ray System Model: 728143
Hazard: Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down
August 9, 2023Medical DevicesFDA
Megadyne Medical Products, Inc. recalls MEGADYNE Suction Coagulators, Product Code 004125 (Suction Coagulator, Handswitching, 10 Fr, 6 inch (15.24cm)), Product
Hazard: An issue with the Suction Coagulator product family may lead to fluid ingress into the handpiece. Fluid ingress can lead to intermittent device activation, non-activation, or self-activation when plugged into the electrosurgery unit (ESU).