RecallWatchUSA

Home / Recalls / Medical Devices

3M Healthcare Business recalls 3M Steri-Drape, Small Drape with Incise Film, REF 1021NS, general surgery drape

Recalled November 30, 2022 · FDA recall Z-0256-2023 · 3M Healthcare Business

Hazard
During a recent investigation, 3M confirmed the liner on the adhesive component of the affected drapes is difficult to remove without damaging the product and may render the product unusable. Additionally, 3M has observed an increase in reported adhesive related skin injuries for these affected lots. · FDA Class II
Units
104,000 units
Sold at
Worldwide distribution.

Description

During a recent investigation, 3M confirmed the liner on the adhesive component of the affected drapes is difficult to remove without damaging the product and may render the product unusable. Additionally, 3M has observed an increase in reported adhesive related skin injuries for these affected lots.

Product
3M Steri-Drape, Small Drape with Incise Film, REF 1021NS, general surgery drape
Manufacturer(s)
3M Healthcare Business

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

3M Healthcare Business recalls 3M Steri-Drape, Small Drape with Incise Film, REF 1021NS, general surgery drape | RecallWatchUSA