RecallWatchUSA

Home / Recalls / Medical Devices

Abbott Diabetes Care, Inc. recalls FreeStyle Flash Blood Glucose Monitors System. Meter made in China; Distributed by TheraSense, Alameda, CA The

Recalled March 26, 2014 · FDA recall Z-1212-2014 · Abbott Diabetes Care

Hazard
Abbott Diabetes Care has identified through internal testing and investigation that all non-applied voltage legacy meters have the potential to produce out of range control solution results and/or erroneously low blood glucose results when used in conjunction with Free Style test strips lots within expiry. · FDA Class I
Units
25,203 units
Sold at
Worldwide Distribution: US (nationwide) and countries of: Austria, Canada, Germany, Israel, Netherlands, Norway, Sweden, Switzerland, United Kingdom.

Description

Abbott Diabetes Care has identified through internal testing and investigation that all non-applied voltage legacy meters have the potential to produce out of range control solution results and/or erroneously low blood glucose results when used in conjunction with Free Style test strips lots within expiry.

Product
FreeStyle Flash Blood Glucose Monitors System. Meter made in China; Distributed by TheraSense, Alameda, CA The FreeStyle Flash meter should only be used with FreeStyle Test Strips and FreeStyle Control Solution. Using other brands of test strips and control solutions with the FreeStyle Meter can product inaccurate results. Intended to monitor blood glucose from samples taken from the body. IVD use only,
Manufacturer(s)
Abbott Diabetes Care, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov