RecallWatchUSA

Home / Recalls / Medical Devices

Abbott Ireland Diagnostics Division recalls ARCHITECT Total T3 Reagent Kit (7K64) consists of: - 1 or 4 Bottle(s) (6.6 mL/27.0 mL) anti-T3 (sheep) coated microp

Recalled December 3, 2014 · FDA recall Z-0474-2015 · Abbott Ireland Diagnostics Division

Hazard
17% of ARCHITECT Total T3 Reagent Kit, LN 7K64-20/-25 reagent lots 38901UI00 and 38901UI01 may exhibit lower Relative Light Units (RLUs) than expected, which in turn may result in controls out of range or patient results higher than expected. · FDA Class II
Units
7,415 kits
Sold at
Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AL , AR , AZ , CA , CT , FL , GA , IL , IN , KS , LA , MA , MD , MI , MN , MO , MS , MT , NC , NJ , NV , NY , OH , OK , PA , SC , SD , TN , TX , UT , VA , WA , WI and WV., and the countries of : Andorra , Angola , Argentina , Armenia , Austria , Azerbaijan , Bangladesh , Barbados , Belarus , Belgium , Bosnia & Herzegovina , Brazil , Burkina Faso , Canada , Cayman Islands , Chile , China , Colombia , Costa Rica , Croatia , Czech Republic , Ecuador , Egypt , El Salvador , Ethopia , Georgia , Germany , Greece , Greenland , Guatemala , Honduras , Hong Kong , India , Indonesia , Iran , Ireland , Israel , Italy , Jamaica , Japan , Kazakhstan , Kenya , Korea , Kuwait , Latvia , Lebanon , Libya , Malaysia , Mexico , Moldova , Montenegro , Oman , Pakistan , Gaza & Jericho , Panama , Paraguay , Peru , Philippines , Poland , Portugal , Qatar , Romania , Russia , Saudi Arabia , Serbia , Singapore , Slovakia , Spain , Sri Lanka. , Switzerland , Taiwan , Tajikistan , Thailand , The Netherlands , Trinidad & Tobago , Turkey , Ukraine ,United Arab Emirates , United Kingdom , Venezuela , Vietnam and Yemen.

Description

17% of ARCHITECT Total T3 Reagent Kit, LN 7K64-20/-25 reagent lots 38901UI00 and 38901UI01 may exhibit lower Relative Light Units (RLUs) than expected, which in turn may result in controls out of range or patient results higher than expected.

Product
ARCHITECT Total T3 Reagent Kit (7K64) consists of: - 1 or 4 Bottle(s) (6.6 mL/27.0 mL) anti-T3 (sheep) coated microparticles in MES buffer with sheep IgG stabilizers. Minimum Concentration: 0.08% solids. Preservative: ProClin 300. - 1 or 4 Bottle(s) (5.9 mL/26.3 mL) T3 acridinium-labeled conjugate in citrate buffer with NaCl and Triton X-100 stabilizers. Minimum concentration: 0.33 ng/mL. Preservative: ProClin 300. The ARCHITECT Total T3 (TT3) assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of total triiodothyronine (Total T33) in human serum and plasma.
Manufacturer(s)
Abbott Ireland Diagnostics Division
Made in
Ireland

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Abbott Ireland Diagnostics Division recalls ARCHITECT Total T3 Reagent Kit (7K64) consists of: - 1 or 4 Bottle(s) (6.6 mL/27.0 mL) anti-T3 (sheep) coated microp | RecallWatchUSA