RecallWatchUSA

Home / Recalls / Medical Devices

Abbott Laboratories recalls Accelerator APS Aliquoter Module, List number: 07L40 - Product Usage: The ACCELERATOR APS Workcell is a modular system d

Recalled June 10, 2020 · FDA recall Z-2251-2020 · Abbott Laboratories

Hazard
In the case of an AQM clot detection error a portion of the distilled water from the Aliquoter Module hydraulic circuit may be dispensed into the Primary Tube leading to sample dilution, which could lead to incorrect results. · FDA Class II
Units
11
Sold at
Worldwide distribution - U.S. Nationwide distribution including in the states of NC, ND, SC, TN. The countries of United Kingdom, France, New Zealand, Pakistan, Turkey.

Description

In the case of an AQM clot detection error a portion of the distilled water from the Aliquoter Module hydraulic circuit may be dispensed into the Primary Tube leading to sample dilution, which could lead to incorrect results.

Product
Accelerator APS Aliquoter Module, List number: 07L40 - Product Usage: The ACCELERATOR APS Workcell is a modular system designed to automate sample handling and processing in the clinical laboratory. The system allows consolidations of multiple clinical chemistry and immunoassay analytical instruments into a unified workstation.
Manufacturer(s)
Abbott Laboratories

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Abbott Laboratories recalls Accelerator APS Aliquoter Module, List number: 07L40 - Product Usage: The ACCELERATOR APS Workcell is a modular system d | RecallWatchUSA