Home / Recalls / Medical Devices
Abbott Laboratories recalls ARCHITECT c4000 REF 02P24-01/2P24, 1P86, 1R24, and 1R25; ARCHITECT c8000 REF 01G06-11/1G06; ARCHITECT c16000 REF 03L77-0
Recalled December 1, 2021 · FDA recall Z-0272-2022 · Abbott Laboratories
Description
Twelve software-related issues affecting software version 9.41 and earlier were identified. The issues include bypassing error codes, incorrect calibration, loss of module communication leading to overfilling wash buffer containers, missing QC flags, signal spikes, abnormal optics reads, and incorrect sample aspiration. The issues may lead to inaccurate results with an unknown bias direction and magnitude for assays measuring cardiac, metabolic (including diabetes), toxicology, prenatal, therapeutic drug monitoring, infectious disease, oncology, hepatic, pancreatic, hematological, endocrinologic, inflammatory analytes.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →