Abbott Laboratories recalls ARCHITECT i1000SR REF 01L86-01/1L86 and 1L87; ARCHITECT i2000SR REF 03M74-02/3M74; ARCHITECT i2000 REF 08C89-01/1G17 and

Recalled December 1, 2021 · FDA recall Z-0271-2022 · Abbott Laboratories

Hazard
Twelve software-related issues affecting software version 9.41 and earlier were identified. The issues include bypassing error codes, incorrect calibration, loss of module communication leading to overfilling wash buffer containers, missing QC flags, signal spikes, abnormal optics reads, and incorrect sample aspiration. The issues may lead to inaccurate results with an unknown bias direction and magnitude for assays measuring cardiac, metabolic (including diabetes), toxicology, prenatal, therapeutic drug monitoring, infectious disease, oncology, hepatic, pancreatic, hematological, endocrinologic, inflammatory analytes. · FDA Class II
Units
20,994 systems
Sold at
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: AFGHANISTAN ALBANIA, ALGERIA, ANDORRA, ANGOLA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHAMAS, BAHRAIN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BELIZE, BENIN, BERMUDA, BOLIVIA, BOSNIA AND HERZEGOVINA, BOTSWANA, BRAZIL, BRUNEI, BULGARIA, BURKINA FASO, BURUNDI, CAMBODIA, CAMEROON, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DEMOCRATIC REP. OF CONGO, DENMARK, FAROE ISLANDS, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ERITREA, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GEORGIA, GERMANY, GHANA, GREECE, GREENLAND, GUATEMALA, GUINEA, HAITI, HONDURAS, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, IVORY COAST, JAMAICA, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KUWAIT, KYRGYZSTAN, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MADAGASCAR, MALAWI, MALAYSIA, MALDIVES, MALI, MALTA, MAURITANIA, MAURITIUS, MEXICO, MOLDOVA, MONTENEGRO, MONTSERRAT, MOROCCO, MOZAMBIQUE, MYANMAR, NAMIBIA, NEPAL, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NICARAGUA, NIGERIA, NORWAY, OMAN, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, RWANDA, SAUDI ARABIA, SENEGAL, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, ST. LUCIA, ST. VINCENT, SURINAME, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TAJIKISTAN, TANZANIA, THAILAND, TRINIDAD & TOBAGO, TUNISIA, TURK & CAICOS, TURKEY, UAE, UGANDA, UK, UKRAINE, URUGUAY, UZBEKISTAN, VIETNAM, WEST BANK/GAZA STRIP, YEMEN, ZAMBIA, ZIMBABWE,

Description

Twelve software-related issues affecting software version 9.41 and earlier were identified. The issues include bypassing error codes, incorrect calibration, loss of module communication leading to overfilling wash buffer containers, missing QC flags, signal spikes, abnormal optics reads, and incorrect sample aspiration. The issues may lead to inaccurate results with an unknown bias direction and magnitude for assays measuring cardiac, metabolic (including diabetes), toxicology, prenatal, therapeutic drug monitoring, infectious disease, oncology, hepatic, pancreatic, hematological, endocrinologic, inflammatory analytes.

Product
ARCHITECT i1000SR REF 01L86-01/1L86 and 1L87; ARCHITECT i2000SR REF 03M74-02/3M74; ARCHITECT i2000 REF 08C89-01/1G17 and 8C89
Manufacturer(s)
Abbott Laboratories