Home / Recalls / Medical Devices
Abbott Laboratories recalls ARCHITECT Lactic Acid Part # 09D891T21 The Lactic Acid assay is used for the quantitation of lactic acid in human pla
Recalled November 16, 2016 · FDA recall Z-0379-2017 · Abbott Laboratories
Hazard
Abbott has identified negative interference from the drug N-Acetyl Cysteine (NAC) with the ARCHITECT Lactic Acid reagent (LN 9D89-21). · FDA Class II
Units
26, 837 kits
Sold at
Nationwide and Worldwide Distribution.
Description
Abbott has identified negative interference from the drug N-Acetyl Cysteine (NAC) with the ARCHITECT Lactic Acid reagent (LN 9D89-21).
Product
ARCHITECT Lactic Acid Part # 09D891T21 The Lactic Acid assay is used for the quantitation of lactic acid in human plasma.
Manufacturer(s)
Abbott Laboratories
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →