Abbott Laboratories recalls Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number 09H39-01. The firm

Recalled October 30, 2019 · FDA recall Z-0173-2020 · Abbott Laboratories

Hazard
The diluent tubing used on the instrument appears cloudy or to have a powder/film on the inner and outer surfaces which may result in falsely elevated platelet results. · FDA Class II
Units
22 analyzers with tubing
Sold at
Distribution was made to GA, LA, MA, NC, NY, SC, and TX. There was no military/government distribution. Foreign distribution was made to Australia, Germany, Greece, Honduras, Indonesia, Italy, and Serbia.

Description

The diluent tubing used on the instrument appears cloudy or to have a powder/film on the inner and outer surfaces which may result in falsely elevated platelet results.

Product
Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number 09H39-01. The firm name on the label is Abbott Laboratories, Abbott Park, IL.
Manufacturer(s)
Abbott Laboratories

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov