Abbott Laboratories, Inc recalls Alinity s System software version2.8.0, REF LN 06P16-01, IVD, OEI: (01)00380740138479 The Alinity s System is intende

Recalled March 23, 2022 · FDA recall Z-0749-2022 · Abbott Laboratories

Hazard
Software error associated with the immunoassay analyzer wash cycle which is using 1 mL of wash buffer instead of the intended 3 mL of wash buffer to wash the exterior of the probe. · FDA Class II
Units
306 units
Sold at
Worldwide distribution: US (nationwide) including states of: AL, CA, GA, IL, IN, MN, MO, NC, OK, PA, RI, SC, TN, TX, WA, WI, MD and countries (OUS) including of: Australia, Austria, Bahrain, Belgium, Brazil, china, Colombia, Croatia, Cyprus, Czech Republic, Egypt, Germany, Greece, Hong Kong, Hungary, Iran, Ireland, Israel, Italy, Jordan Kuwait and Mexico.

Description

Software error associated with the immunoassay analyzer wash cycle which is using 1 mL of wash buffer instead of the intended 3 mL of wash buffer to wash the exterior of the probe.

Product
Alinity s System software version2.8.0, REF LN 06P16-01, IVD, OEI: (01)00380740138479 The Alinity s System is intended for In Vitro diagnostic use only. The Alinity s System is a highthroughput, fully automated immunoassay analyzer designed to determine the presence of specific antigens and antibodies by using chemiluminescent immunoassay technology.
Manufacturer(s)
Abbott Laboratories, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov