Abbott Laboratories, Inc recalls ARCHITECT MAGNESIUM, (a) REF 7D70-21, (b) 7D70-31 The Magnesium assay is used for the quantitation of magnesium in hu

Recalled March 13, 2019 · FDA recall Z-0971-2019 · Abbott Laboratories

Hazard
Abbott internal testing has identified that the Magnesium urine application demonstrates depressed urine result recovery. All samples >0.5mEq/L fail the linearity acceptance criteria for the bias specification and demonstrate depressed recovery of up to 37%. The specific cause of the negative bias/depressed results is currently under investigation. · FDA Class II
Units
24,995 units
Sold at
Worldwide Distribution: US (nationwide) and to countries of: CHINA, THAILAND, GUATEMALA, and EL SALVADOR.

Description

Abbott internal testing has identified that the Magnesium urine application demonstrates depressed urine result recovery. All samples >0.5mEq/L fail the linearity acceptance criteria for the bias specification and demonstrate depressed recovery of up to 37%. The specific cause of the negative bias/depressed results is currently under investigation.

Product
ARCHITECT MAGNESIUM, (a) REF 7D70-21, (b) 7D70-31 The Magnesium assay is used for the quantitation of magnesium in human serum, plasma, or urine.
Manufacturer(s)
Abbott Laboratories, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov