RecallWatchUSA

Home / Recalls / Medical Devices

Abbott Laboratories, Inc recalls Clinical Chemistry Phenobarbital, list number 1E08

Recalled April 24, 2013 · FDA recall Z-1118-2013 · Abbott Laboratories

Hazard
Abbott Diagnostics is recalling Clinical Chemistry Phenobarbital reagent lots 62299UN12 and 85773UN12 due to imprecision issues which increase with the age of the reagent lot. The imprecision of the Phenobarbital assay is d 7% Total CV%; however, internal testing confirmed CVs up to approximately 12% for in-date reagent lots over five months old. Although lot 52803UN12 is now expired, it exhibit · FDA Class II
Units
1289
Sold at
Worldwide Distribution - USA (nationwide) and Internationally to Internationally distributed to Argentina, Australia, Austria, Bahamas, Brunei, Canada, Chile, Curacao, Denmark, France, Germany, Greece, Ireland, Italy, Lebanon, Mexico, Netherlands, Norway, Oman, Paraguay, Portugal, Puerto Rico, Russia, Saudi Arabia, Serbia, South Korea, Spain, Switzerland, Turkey, and United Kingdom.

Description

Abbott Diagnostics is recalling Clinical Chemistry Phenobarbital reagent lots 62299UN12 and 85773UN12 due to imprecision issues which increase with the age of the reagent lot. The imprecision of the Phenobarbital assay is d 7% Total CV%; however, internal testing confirmed CVs up to approximately 12% for in-date reagent lots over five months old. Although lot 52803UN12 is now expired, it exhibit

Product
Clinical Chemistry Phenobarbital, list number 1E08
Manufacturer(s)
Abbott Laboratories, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Abbott Laboratories, Inc recalls Clinical Chemistry Phenobarbital, list number 1E08 | RecallWatchUSA