RecallWatchUSA

Home / Recalls / Medical Devices

Abbott Laboratories Inc. (St Jude Medical) recalls TactiSys Quartz Equipment (with software version 1.7) Model: PN-004 400 Part Numbers: 600043767 (China), 100154367 (

Recalled June 5, 2019 · FDA recall Z-1493-2019 · Abbott Laboratories Inc. (St Jude Medical)

Hazard
In reported cases, the device log on the TactiSys Quartz Equipment operating on Software Version 1.7.0 fills the allocated disk space, which prevents the storage of new log data. This may lead to intermittent contact force data to be displayed during the procedure. · FDA Class II
Units
317 units
Sold at
US, UNITED KINGDOM, UAE, TURKEY, THAILAND, TAIWAN, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVENIA, SLOVAKIA, SINGAPORE, SERBIA, SAUDIA ARABIA, RUSSIA, ROMANIA, PORTUGAL, POLAND, PALESTINE, OMAN, NORWAY, NEW ZEALAND, NETHERLANDS, MOROCCO, MALAYSIA, LEBANON, JORDAN, JAPAN, ITALY, ISRAEL, IRELAND, IRAN, GREECE, GERMANY, FRANCE, EL SALVADOR, EGYPT, DENMARK, Czech Republic, CHINA, CANADA, BULGARIA, BELGIUM, AUSTRIA, AUSTRALIA

Description

In reported cases, the device log on the TactiSys Quartz Equipment operating on Software Version 1.7.0 fills the allocated disk space, which prevents the storage of new log data. This may lead to intermittent contact force data to be displayed during the procedure.

Product
TactiSys Quartz Equipment (with software version 1.7) Model: PN-004 400 Part Numbers: 600043767 (China), 100154367 (Global), 100154368 (SWAP)
Manufacturer(s)
Abbott Laboratories Inc. (St Jude Medical)

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Abbott Laboratories Inc. (St Jude Medical) recalls TactiSys Quartz Equipment (with software version 1.7) Model: PN-004 400 Part Numbers: 600043767 (China), 100154367 ( | RecallWatchUSA