RecallWatchUSA

Home / Recalls / Medical Devices

Abbott Medical Optics, Inc. recalls STAR Excimer Laser System, Model STAR S4 IR, Catalog Number: 0030-4864, Product is manufactured and distributed by AMO M

Recalled October 30, 2013 · FDA recall Z-0049-2014 · Abbott Medical Optics

Hazard
This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the AC Power Distribution Printed Circuit Board (PCB). · FDA Class II
Units
834 units in total
Sold at
Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Asia Pacific, Canada, Europe, Africa, Middle East, Japan, and Latin America.

Description

This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the AC Power Distribution Printed Circuit Board (PCB).

Product
STAR Excimer Laser System, Model STAR S4 IR, Catalog Number: 0030-4864, Product is manufactured and distributed by AMO Manufacturing USA, LLC Milpitas, CA The STAR Excimer Laser Systems are indicated for laser assisted in situ keratomileusis (LASIK), photorefractive keratectomy (PRK), and phototherapeutic keratectomy (PTK) eye surgery in certain patient populations.
Manufacturer(s)
Abbott Medical Optics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Abbott Medical Optics, Inc. recalls STAR Excimer Laser System, Model STAR S4 IR, Catalog Number: 0030-4864, Product is manufactured and distributed by AMO M | RecallWatchUSA