Abbott recalls Thoratec HeartMate 3 LVAS Implant Kit, Rx Only, Catalog # 106524US, 106524, 10652INT - LVAS KIT, HM 3 Indicated for p

Recalled May 23, 2018 · FDA recall Z-1774-2018 · Abbott

Hazard
Reports of outflow graft twist occlusions. Patients whose devices experience outflow graft occlusions will experience a persistent low flow alarm. Outflow graft occlusions can result in serious adverse events such as hemodynamic compromise, thrombus, and death. · FDA Class I
Units
5,607
Sold at
Worldwide Distribution -- U.S., including the states of MN, PA, IL, WI, AZ, AR, TN, MO, TX, MA, NE, MI, CA, NC, OH, NY, GA, VA, FL, OK, MD, ME, SC, IN, LA, OR, WA, AL, CO, IA, KS, KY, UT, DC, and CT; and, the countries of Australia, Austria, Belgium, Brazil, Brunei, Canada, Cayman Islands, Colombia, Croatia, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Italy, Kuwait, Lebanon, Malaysia, Montenegro, Netherlands, Poland, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.

Description

Reports of outflow graft twist occlusions. Patients whose devices experience outflow graft occlusions will experience a persistent low flow alarm. Outflow graft occlusions can result in serious adverse events such as hemodynamic compromise, thrombus, and death.

Product
Thoratec HeartMate 3 LVAS Implant Kit, Rx Only, Catalog # 106524US, 106524, 10652INT - LVAS KIT, HM 3 Indicated for providing short-term hemodynamic support in patients with advanced refractory left ventricular heart failure.
Manufacturer(s)
Abbott

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Abbott recalls Thoratec HeartMate 3 LVAS Implant Kit, Rx Only, Catalog # 106524US, 106524, 10652INT - LVAS KIT, HM 3 Indicated for p | RecallWatchUSA