RecallWatchUSA

Home / Recalls / Medical Devices

Abbott Vascular recalls Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only

Recalled June 5, 2019 · FDA recall Z-1687-2019 · Abbott Vascular

Hazard
Reports of implantable mitral valve repair system clips unexpectedly opening and becoming nonfunctional, resulting from unintended excessive force applied during the Clip implantation. Excessive force can also result in unexpected movement of clip arms. The inability to close and remove the device has lead to surgery and additional intervention. · FDA Class II
Units
N/A
Sold at
U.S.: FL, UT, IN, OK, TX, CO, LA, AZ, MO, AL, GA, CA, AR, KS, IL, HI, IA, KY, MA, DC, NY, NV, NM, NJ, MT, MD, NC, CT, MI, MN, OH, ND, NH, OR, PA, WI, WA, VA, WV, TN, SC, RI, DE, ME, NE, SD, MS, ID, VT. Foreign (OUS): AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, Canada, CROATIA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, INDONESIA, IRELAND, ISRAEL, ITALY, KUWAIT, LEBANON, MACAU, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UNITED KINGDOM

Description

Reports of implantable mitral valve repair system clips unexpectedly opening and becoming nonfunctional, resulting from unintended excessive force applied during the Clip implantation. Excessive force can also result in unexpected movement of clip arms. The inability to close and remove the device has lead to surgery and additional intervention.

Product
Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only
Manufacturer(s)
Abbott Vascular

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Abbott Vascular recalls Abbott MitraClip XTR Clip Delivery System, UDI: 08717648226366, Part: CDS0601-XTR, Sterile EO, Rx Only | RecallWatchUSA