Abiomed, Inc. recalls lmpella CP with Smart Assist, Model Number: 0048-0003. Left heart support blood pump, for Use During Cardiogenic Shock

Recalled November 13, 2019 · FDA recall Z-0345-2020 · Abiomed

Hazard
This one pump shipped with incorrect pump programming, the initial production working version of the program. The pump would have immediately run at P9 for all selected P-levels, rather than a 10 second ramp up period. The pump may have migrated out of position and possibly caused injury. The positioning and suction alarms were also disabled. · FDA Class II
Units
1
Sold at
The product was distributed to GA

Description

This one pump shipped with incorrect pump programming, the initial production working version of the program. The pump would have immediately run at P9 for all selected P-levels, rather than a 10 second ramp up period. The pump may have migrated out of position and possibly caused injury. The positioning and suction alarms were also disabled.

Product
lmpella CP with Smart Assist, Model Number: 0048-0003. Left heart support blood pump, for Use During Cardiogenic Shock and High-Risk PCI
Manufacturer(s)
Abiomed, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Abiomed, Inc. recalls lmpella CP with Smart Assist, Model Number: 0048-0003. Left heart support blood pump, for Use During Cardiogenic Shock | RecallWatchUSA