RecallWatchUSA

Home / Recalls / Medical Devices

Accriva Diagnostics, Inc. recalls AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter.

Recalled November 6, 2024 · FDA recall Z-0279-2025 · Accriva Diagnostics

Hazard
Values used to calibrate whole blood oximeter instruments were observed to be higher than normal, which may result in incorrect patient results where oxygen levels are higher than actual results, which may lead to patient care mismanagement requiring medical intervention. · FDA Class II
Units
1
Sold at
US: AZ

Description

Values used to calibrate whole blood oximeter instruments were observed to be higher than normal, which may result in incorrect patient results where oxygen levels are higher than actual results, which may lead to patient care mismanagement requiring medical intervention.

Product
AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter.
Manufacturer(s)
Accriva Diagnostics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Accriva Diagnostics, Inc. recalls AVOXimeter 1000E, Product Number: AVOX1000E. The AVOXimeter 1000E is a battery-operated bedside whole blood oximeter. | RecallWatchUSA