RecallWatchUSA

Home / Recalls / Medical Devices

Accuray Incorporated recalls Ashland(TM) Box Film Ball Cube II EBT3, P/N: 027863-002 rev E Product is used to calibrate CyberKnife systems

Recalled September 26, 2018 · FDA recall Z-3211-2018 · Accuray Incorporated

Hazard
the Ball Cube II phantom calibration film was not manufactured within necessary dimensions and may introduce up to 0.5 mm of positional inaccuracy to the CyberKnife System · FDA Class II
Units
259 units
Sold at
Worldwide Distribution: United States (nationwide), and countries of: United Kingdom, Germany, France, Finland, Netherlands, Portugal, Hong Kong, Poland, Australia, United Arab Emirates, Thailand, and Japan.

Description

the Ball Cube II phantom calibration film was not manufactured within necessary dimensions and may introduce up to 0.5 mm of positional inaccuracy to the CyberKnife System

Product
Ashland(TM) Box Film Ball Cube II EBT3, P/N: 027863-002 rev E Product is used to calibrate CyberKnife systems
Manufacturer(s)
Accuray Incorporated

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov