Acist Medical Systems recalls ACIST Kodama Intravascular Ultrasound Catheter The Kodama Intravascular Ultrasound Catheter is a component of the AC
Recalled March 10, 2021 · FDA recall Z-1161-2021 · Acist Medical Systems
Hazard
Test results from the manufacturing line found a piece of damaged o-ring in an unexpected section of the catheter. Further testing indicated that pieces (>200 micron) of damaged o-ring had the potential to be flushed out of the catheter. ACIST is confirming the source of the failure mode to assure the quality and reliability of the Kodama catheter. There have been no related field reports related to this incident, nor any evidence or report of patient injury or adverse health consequence. · FDA Class I
Units
1185 units
Sold at
Worldwide - US Nationwide Distribution in the states of AL, AR, CA, CO, FL, IL, KS, KY, LA, MD, MI, MO, NC, NJ, NY, OK, PA, RI, TN; In the countries of India, Italy, Japan, Poland, and United Arab Emirates.
Description
Test results from the manufacturing line found a piece of damaged o-ring in an unexpected section of the catheter. Further testing indicated that pieces (>200 micron) of damaged o-ring had the potential to be flushed out of the catheter. ACIST is confirming the source of the failure mode to assure the quality and reliability of the Kodama catheter. There have been no related field reports related to this incident, nor any evidence or report of patient injury or adverse health consequence.
Product
ACIST Kodama Intravascular Ultrasound Catheter The Kodama Intravascular Ultrasound Catheter is a component of the ACIST HDi System. The ACIST HDi System is intended to be used for ultrasound examination of coronary and peripheral intravascular pathology. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures. The ACIST Kodama Intravascular Ultrasound Catheter is intended for use with the ACIST HDi System.
Manufacturer(s)
Acist Medical Systems