Ad-Tech Medical Instrument Corporation recalls Ad-Tech Medical Instrument Corporation Macro Micro Subdural Electrode. For temporary (<30 days) use with recording, mon
Recalled March 13, 2013 · FDA recall Z-0892-2013 · Ad-Tech Medical Instrument
Hazard
The microelectrodes are not flush with the silastic substrate, may contain sharp edges capable of abrading the cortical surface, and may have remaining pieces which may fracture and remain in the cortical tissue with the electrode is explanted. Recall expanded on 5/2/2013 to include additional Macro Micro Subdural Electrodes. · FDA Class I
Units
144 (103 US; 41 OUS)
Sold at
Worldwide Distribution - USA Nationwide including the states of CA, GA, LA, MA, MD, MI, MN, NH, PA, RI, TX, UT, WI, NY IA, WI and the countries of: Australia, CANADA, ENGLAND, JAPAN, KOREA, SPAIN, FRANCE, and SWITZERLAND.
Description
The microelectrodes are not flush with the silastic substrate, may contain sharp edges capable of abrading the cortical surface, and may have remaining pieces which may fracture and remain in the cortical tissue with the electrode is explanted. Recall expanded on 5/2/2013 to include additional Macro Micro Subdural Electrodes.
Product
Ad-Tech Medical Instrument Corporation Macro Micro Subdural Electrode. For temporary (<30 days) use with recording, monitoring and stimulation equipment, for the recording, monitoring and stimulation of electrical signals on the surface level of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain mapping.
Manufacturer(s)
Ad-Tech Medical Instrument Corporation