Ad-Tech Medical Instrument Corporation recalls AD-TECH Spencer Probe Depth Electrode, REF SD04R-AP58X-000 4-Contact 1.12mm (A style) Depth Electrode

Recalled October 30, 2024 · FDA recall Z-0219-2025 · Ad-Tech Medical Instrument

Hazard
Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above. · FDA Class II
Units
2 units
Sold at
US Nationwide distribution in the states of CA, OH.

Description

Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.

Product
AD-TECH Spencer Probe Depth Electrode, REF SD04R-AP58X-000 4-Contact 1.12mm (A style) Depth Electrode
Manufacturer(s)
Ad-Tech Medical Instrument Corporation