RecallWatchUSA

Home / Recalls / Medical Devices

Ad-Tech Medical Instrument Corporation recalls Disposable Drill Kit which includes two drill bits, two drill stops, and two adjustment wrenches, sterile. The drill bit

Recalled June 6, 2018 · FDA recall Z-1938-2018 · Ad-Tech Medical Instrument

Hazard
There is a possibility that DDK2-2.4-30X Disposable Drill Kits, Lot Number 111664 208140649 contained 2.8mm drill bits from DDK2-2.8-30X Disposable Drill Kits, Lot Number 111745 208140649 and vice versa. · FDA Class II
Units
30 devices
Sold at
Worldwide Distribution: US (nationwide) to states of: CA, DC, FL, MI, MA, NC, NY, WA, WI; and country of: Canada. Distributed internationally to the United Kingdom.

Description

There is a possibility that DDK2-2.4-30X Disposable Drill Kits, Lot Number 111664 208140649 contained 2.8mm drill bits from DDK2-2.8-30X Disposable Drill Kits, Lot Number 111745 208140649 and vice versa.

Product
Disposable Drill Kit which includes two drill bits, two drill stops, and two adjustment wrenches, sterile. The drill bits are 30cm long and made of stainless steel. The drill kits are supplied in a sterile state and are single-use only. Used to drill cranial holes using a stereotactic frame.
Manufacturer(s)
Ad-Tech Medical Instrument Corporation