RecallWatchUSA

Home / Recalls / Medical Devices

Advanced Bionics, LLC recalls HiResolution Bionic Ear System HiRes 90K Advantage CI HiFocus MS Electrode REF CI-1500-04. Cochlear Implant.

Recalled February 26, 2020 · FDA recall Z-1259-2020 · Advanced Bionics

Hazard
Three materials used in the makeup of the implant were mistakenly left off the list of bio-compatible materials in the firm's previously approved submission to the Pharmaceuticals and Medical Devices Agency (PMDA) of Japan. · FDA Class III
Units
4
Sold at
No US distribution OUS: Japan

Description

Three materials used in the makeup of the implant were mistakenly left off the list of bio-compatible materials in the firm's previously approved submission to the Pharmaceuticals and Medical Devices Agency (PMDA) of Japan.

Product
HiResolution Bionic Ear System HiRes 90K Advantage CI HiFocus MS Electrode REF CI-1500-04. Cochlear Implant.
Manufacturer(s)
Advanced Bionics, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov