Aesculap Inc recalls Aeos Robotic Digital Microscope, Product Code: PV010

Recalled June 19, 2024 · FDA recall Z-2139-2024 · Aesculap

Hazard
Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism. · FDA Class II
Units
19 units
Sold at
US Nationwide distribution in the states of AR, FL, IL, IN, KY, MI, NC, NE, OK, SD, TX.

Description

Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.

Product
Aeos Robotic Digital Microscope, Product Code: PV010
Manufacturer(s)
Aesculap Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov