RecallWatchUSA

Home / Recalls / Medical Devices

Aesculap, Inc. recalls Valve XS Atrium Retractor FC429R 52154756 The Valve XS Atrium Lift Retractor is used in heart surgery. The retracto

Recalled January 6, 2016 · FDA recall Z-0566-2016 · Aesculap

Hazard
AAG received a customer complaint that corrosion had been found by a hospitals central sterile service department on the ball joint of an FC424R blade. Investigation revealed that the corrosion was due to incorrect material used in the manufacturing of the disk spring of the ball joint by the supplier · FDA Class II
Units
41
Sold at
US Distribution to state of: NY.

Description

AAG received a customer complaint that corrosion had been found by a hospitals central sterile service department on the ball joint of an FC424R blade. Investigation revealed that the corrosion was due to incorrect material used in the manufacturing of the disk spring of the ball joint by the supplier

Product
Valve XS Atrium Retractor FC429R 52154756 The Valve XS Atrium Lift Retractor is used in heart surgery. The retractor is used for the temporary dilation of the left atrium in mitral valve surgery and for assistance in tricuspid valve surgery.
Manufacturer(s)
Aesculap, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Aesculap, Inc. recalls Valve XS Atrium Retractor FC429R 52154756 The Valve XS Atrium Lift Retractor is used in heart surgery. The retracto | RecallWatchUSA