RecallWatchUSA

Home / Recalls / Medical Devices

Aizu Olympus Co., Ltd. recalls Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endosc

Recalled November 29, 2023 · FDA recall Z-0328-2024 · Aizu Olympus

Hazard
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection · FDA Class II
Units
823 units
Sold at
Worldwide - US Nationwide distribution.

Description

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Product
Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the airways, trancheobronchial tree, esophagus and surrounding organs Model Numbers: BF-UC180F, BF-UC190F.
Manufacturer(s)
Aizu Olympus Co., Ltd.
Made in
Japan

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov