Aizu Olympus Co., Ltd. recalls Colonoscope: for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid
Recalled November 29, 2023 · FDA recall Z-0329-2024 · Aizu Olympus
Hazard
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection · FDA Class II
Units
3344 units
Sold at
Worldwide - US Nationwide distribution.
Description
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Product
Colonoscope: for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve). Model Numbers: CF-2T160I, CF-2T160L, CF-H170I, CF-H170L, CF-H180AI, CF-H180AL, CF-H180DI, CF-H180DL, CF-H190I, CF-H190L, CF-HQ190I, CF-HQ190L, CF-LV1I, CF-LV1L, CF-Q150I, CF-Q150L, CF-Q160S, CF-Q180AI, CF-Q180AL. 510(k) Numbers: K954451, K112680, K100584, K051645, K111756, K001241, K911278.