RecallWatchUSA

Home / Recalls / Medical Devices

Aizu Olympus Co., Ltd. recalls Cystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-V

Recalled November 29, 2023 · FDA recall Z-0331-2024 · Aizu Olympus

Hazard
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection · FDA Class II
Units
1243 units
Sold at
Worldwide - US Nationwide distribution.

Description

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Product
Cystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.
Manufacturer(s)
Aizu Olympus Co., Ltd.
Made in
Japan

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Aizu Olympus Co., Ltd. recalls Cystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-V | RecallWatchUSA