RecallWatchUSA

Home / Recalls / Medical Devices

Alcon Research, LTD. recalls Alcon 25+¿ TOT ALPLUS¿ Vitrectomv Pak Includes all necessary items for a core vitrectomy procedure and connects to th

Recalled March 7, 2018 · FDA recall Z-0778-2018 · Alcon Research

Hazard
Alcon is conducting this Voluntary Medical Device Recall for a specific lot (2060953H) of its 25+ TOTALPLUS Vitrectomy Pak due a potential that a 23GA ULTRAVIT probe may be included within the 25+ TOTALPLUS Vitrectomy Pak, instead of a 25+ ULTRAVIT probe. · FDA Class II
Units
300 units
Sold at
Worldwide Distribution : Argentina Australia France Italy Mexico Poland Spain Sweden Russia South Korea Ukraine

Description

Alcon is conducting this Voluntary Medical Device Recall for a specific lot (2060953H) of its 25+ TOTALPLUS Vitrectomy Pak due a potential that a 23GA ULTRAVIT probe may be included within the 25+ TOTALPLUS Vitrectomy Pak, instead of a 25+ ULTRAVIT probe.

Product
Alcon 25+¿ TOT ALPLUS¿ Vitrectomv Pak Includes all necessary items for a core vitrectomy procedure and connects to the CONSTELLATION Vision System.
Manufacturer(s)
Alcon Research, LTD.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov