Home / Recalls / Medical Devices
Alcon Research, LTD. recalls CyPass System 241-S, CYPASS MICRO-STENT, CYPASS LOADED, CYPASS APPLIER, REF CATALOG #: 8065754003 For ophthalmic use.
Recalled October 31, 2018 · FDA recall Z-0185-2019 · Alcon Research
Hazard
Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent cataract surgery alone. · FDA Class I
Units
30561 units
Sold at
Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, France, Finland, Germany, Hong Kong, Iceland, Italy, Mexico, Peru, Singapore, Spain, Sweden, Switzerland, United Kingdom, and Uruguay.
Description
Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent cataract surgery alone.
Product
CyPass System 241-S, CYPASS MICRO-STENT, CYPASS LOADED, CYPASS APPLIER, REF CATALOG #: 8065754003 For ophthalmic use.
Manufacturer(s)
Alcon Research, LTD.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →