RecallWatchUSA

Home / Recalls / Medical Devices

Alere San Diego, Inc. recalls Alere INRatio2 PT/INR Professional Test Strips, Model Number: 99008G2. The Alere INRatio2 PT/INR Monitoring System (Prof

Recalled May 14, 2014 · FDA recall Z-1546-2014 · Alere San Diego

Hazard
Test Strips may report an inaccurately low INR result. Several patients had a therapeutic or near-therapeutic INR with the Alere INRatio2 PT/INR Professional Test Strip but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory. · FDA Class I
Units
Estimated 99,795 Total units
Sold at
US Nationwide Distribution

Description

Test Strips may report an inaccurately low INR result. Several patients had a therapeutic or near-therapeutic INR with the Alere INRatio2 PT/INR Professional Test Strip but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Product
Alere INRatio2 PT/INR Professional Test Strips, Model Number: 99008G2. The Alere INRatio2 PT/INR Monitoring System (Professional Use), consisting of the Monitor and test strip, is used for quantitative determination of international normalized ratio (INR) in fresh, capillary whole blood to monitor the effect of warfarin therapy on clotting time by health care professionals. (in vitro diagnostic use).
Manufacturer(s)
Alere San Diego, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Alere San Diego, Inc. recalls Alere INRatio2 PT/INR Professional Test Strips, Model Number: 99008G2. The Alere INRatio2 PT/INR Monitoring System (Prof | RecallWatchUSA