Alere San Diego, Inc. recalls Alere Triage CardioProfiler Panel PN 97100CP Used for the quantitative determination of creatine kinase MB, myoglobin
Recalled July 18, 2012 · FDA recall Z-1928-2012 · Alere San Diego
Hazard
Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing. · FDA Class I
Units
16,962 kits
Sold at
Nationwide Distribution
Description
Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
Product
Alere Triage CardioProfiler Panel PN 97100CP Used for the quantitative determination of creatine kinase MB, myoglobin, troponin I and B-type natriuretic peptide in EDTA anticoagulated whole blood and plasma specimens.
Manufacturer(s)
Alere San Diego, Inc.