Alere San Diego Inc. DBA Immunalysis Corporation recalls lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Controls (High 125 ng/ml and Low 75 ng/ml) are used as control ma
Recalled June 20, 2018 · FDA recall Z-2140-2018 · Alere San Diego Inc.
Hazard
The High (125 ng/ml) Control does not meet specification - it has been confirmed to produce values below 100 ng/mL when run on validated clinical chemistry analyzers, which is below the 100 ng/mL cutoff of the EDDP Specific Urine Enzyme Immunoassay with which this control is utilized. · FDA Class III
Units
37 kits
Sold at
US Distribution to the states of : CA, CO, CT, FL, LA, MD, NC, NJ, OH, OK, TN, TX, WI, and WV
Description
The High (125 ng/ml) Control does not meet specification - it has been confirmed to produce values below 100 ng/mL when run on validated clinical chemistry analyzers, which is below the 100 ng/mL cutoff of the EDDP Specific Urine Enzyme Immunoassay with which this control is utilized.
Product
lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Controls (High 125 ng/ml and Low 75 ng/ml) are used as control materials in the ammonolysis EDDP Specific Urine Enzyme Immunoassay, which has a cutoff of 100 ng/ml. This test provides only a preliminary analytical test result and a more specific alternate chemical method, such as Gas Chromatography/Mass Spectrometry or Liquid Chromatography/Tandem Mass Spectrometry, must be used to obtain a confirmed analytical result.
Manufacturer(s)
Alere San Diego Inc. DBA Immunalysis Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →