RecallWatchUSA

Home / Recalls / Medical Devices

Alfa Wassermann, Inc. recalls Alfa Wassermann System Diluent 750 mL Manufactured for Alfa Wassermann Diagnostic Technologies LLC 4 Henderson Drive

Recalled February 6, 2013 · FDA recall Z-0743-2013 · Alfa Wassermann

Hazard
Inorganic phosphorus assays using System Diluent Lot 155599 may not product acceptable calibration results. · FDA Class II
Units
4,104 kits
Sold at
Worldwide Distribution - USA (nationwide) and the countries of Canada and Cayman Islands

Description

Inorganic phosphorus assays using System Diluent Lot 155599 may not product acceptable calibration results.

Product
Alfa Wassermann System Diluent 750 mL Manufactured for Alfa Wassermann Diagnostic Technologies LLC 4 Henderson Drive West Caldwell, NJ 07006 Alfa Wassermann B.V. Pompmolenlaan 24 3447 GK, Woerden The Netherlands Common name is water solution. Intended to dilute reagents and rinse the parts of a clinical chemistry instrument.
Manufacturer(s)
Alfa Wassermann, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Alfa Wassermann, Inc. recalls Alfa Wassermann System Diluent 750 mL Manufactured for Alfa Wassermann Diagnostic Technologies LLC 4 Henderson Drive | RecallWatchUSA