Hazard
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide. · FDA Class III
Units
60 kits
Sold at
Nationwide Distribution.
Description
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
Product
ProScreen 12 Panel Dip Card, Item No. PSD-12BUP
Manufacturer(s)
Ameditech Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →